תיאור
Technical Details of the optris MS series
Measurement Specifications
- Temperature range:
MS: -32°C to 420°C (-20°F to 788°F)
MSplus: -32°C to 530°C (-20°F to 980°F)
MSpro: -32°C to 760°C (-20°F to 1440°F) - System accuracy (at Tamb = 23 ±5°C):
± 1% or ± 1°C (20…420/530/760°C)
± 1.5°C (19.9…0°C)
± 2.5°C (-0.1…-20°C)
± 3°C (-20.1…-32°C) - Repeatability:
MS: ± 0.5% or ± 0.7°C (20…420°C)
MSplus: ± 0.5% or ± 0.7°C (20…530°C)
MSpro: ± 0.75% or ± 0.75°C (20…760°C) - Optical resolution (D:S):
MS: 20:1, 13 mm spot size (D < 140 mm)
MSplus: 20:1, 13 mm spot size (D < 140 mm)
MSpro: 40:1, 13 mm spot size (D < 260 mm) - Resolution (display):
MS: 0.2°C (0.5°F)
MSplus: 0.1°C (0.1°F)
MSpro: 0.1°C (0.1°F) - Response time (95%): 300 ms
- Ambient temperature: 0°C to 50°C
- Storage temperature: -20°C to 60°C without battery
- Spectral range: 8 to 14 µm
- Emissivity:
MS: fixed: 0.95
MSplus: 0.100 – 1.100 adjustable
MSpro: 0.100 – 1.100 adjustable - Configurations:
MS: Min / Max / Hold / °C / °F
MSplus: Min / Max / Hold / °C / °F / Offset
MSpro: Min / Max / Hold / °C / °F / Offset - Alarm functions:
MS: none
MSplus & MSpro: Visual LCD alarmcolors and acoustic HIGH-/LOW-alarm - PC Interface, software, thermocouple input:
MS: USB interface
MSplus: USB interface, OptrisConnect report software
MSpro: USB interface, optris Connect report software, thermocouple element type K - Laser: <1 mW laser class IIa, laser beam with 9 mm offset
- Weight / dimensions:
MS: 150 g, 190 x 38 x 45 mm
MSplus: 150 g, 190 x 38 x 45 mm
MSpro: 180 g; 190 x 38 x 45 mm - Battery: 9 V alkaline battery
- Battery life:
40 hours with laser and backlight off,
20 hours with laser and backlight on 50%
- Relative humidity: 10 – 95 % RH non condensing, at <30°C ambient temperature
- Scope of Supply:
MS: 9 V alkaline battery, operators manual
MSplus: 9 V alkaline battery, operators manual, wrist strap, pouch, protection boot, adapter for photo tripod, USB interface cable, OptrisConnect Report software
MSpro: 9 V alkaline battery, operators manual, wrist strap, pouch, protection boot, adapter for photo tripod, USB interface cable, OptrisConnect Report software, thermocouple insertion probe type K
Versions
Part number: temperature range / spectral range / optics / output
- OPTMSI: -32-420 °C / 8-14 μm / 20:1
- OPTMSPLI: -32-530 °C / 8-14 μm / 20:1
- OPTMSPRI: -32-760 °C / 8-14 μm / 40:1